The handheld coil FESP 5133-1330 was designed for immunity testing against magnetic fields up to approx. 20 kHz. The coil has 225 turns of Cu-wire on 15 layers. There are spacers to provide a constant distance of 50 mm between coil center and EuT-surface. The generated magnetic fieldstrength is direct proportional to the coil current. Typical Applications are Immunity Tests according to the Measuring Method SF 01 G acc. to VG 95377 standard.
|
Medium coil diameter: |
126 mm |
|
Medium Coil Distance to Surface: |
50 mm |
|
Number of turns: |
225 |
|
Wire Diameter: |
2 mm |
|
Mechanical Dimensions: |
190 x 190 x 290 mm |
|
Terminals: |
4 mm Laborstecker |
|
Usable Frequency Range: |
0 - 20 (50) kHz |
|
Inductance: |
6.5 mH |
|
Resistance: |
0.5 Ω |
|
Capacitance: |
240-300 pF |
|
Resonance Frequency: |
120 kHz |
|
Impedance: |
|Z|=42 Ω @ 1 kHz |
|
Max. Current: |
20 A |
|
Max. Fieldstrength at 50 mmspacing: |
17 kA/m |
|
Max. magnetic fieldstrength at 1 A coil current in 50 mm distance:
|
858.2 A/m |
|
Required Current for 1000 A/m in 50 mm distance: |
1.165 A |
|
Magnetic Flux Density at 1 A coil current in 50 mm distance: |
1.078 mT 180.66 dBpT
|
|
Conversion Current-Magnetic Fielstrength: |
58.67 dB/m |
|
Weight:
|
5.3 kg
|
|
Standard:
|
Measuring Method SF 01 G acc. to VG95377
|
Datasheet FESP 5133-1330
Security Notes FESP 5133-1330
|