Radiating Loops

 

FESP 5139 - Radiating Loop

 
  • FESP 5139

 

The magnetic, handheld coil FESP 5139 was designed to generate defined mag-netic fieldstrength in the frequency range from DC up to 5 (7) MHz. The main application is immunity testing against magnetic fields.

Depending on the current source char-acteristics magnetic fields up to 360 A/m can be generated for a short time. The generated magnetic fieldstrength is pro-portional to the coil current. A distance ring allows a precise spacing of 50 mm between coil and EUT surface. Other distance rings can be used to scale the current-fieldstrength ratio. The highest fieldstrength levels can be achieved using the shortest ring. There are further rings available on request in order to provide a certain scaling between coil current and magnetic fieldstrength (e.g. 1 A coil current = 20 A/m fieldstrength).

 

Application as Sensor for Magnetic Fields:

The FESP 5139 can also be used to measure existing magnetic fields. For this purpose the FESP 5139 is connected to a frequency selective 50 Ohm measuring instrument e.g. a spectrum analyzer or a measuring receiver.

 

Number of turns: 3
Wire diameter: 1 mm Cu
Maximum Coil Current: 12 A (5 min.)
Nominal Coil Current: 8 A continuous
Maximum magnetic Field Strength:(without distance ring, in coil center) 360 A/m (5 min.)
Nominal Magnetic Field Strength with 50 mm distance ring: 85 A/m continuous
Magnetic Fieldstrength, 1 A Coil Current:
(with 50 mm distance ring)

10.61 A/m

Current required for 1 A/m:
(with 50 mm distance ring)

94.25 mA
Medium coil diameter: 100 mm
Spacing coil center to measurement plane: 50 mm
Mechanical Dimensions: 11 x 11 x 21 cm
Connector: N-female
Usable Frequency Range: 0 - 5 (7) MHz
Inductance: 2.3 μH

Weight:

0.47 kg

 

Datasheet FESP 5139

 

 


 

FESP 5133-9 - Circular screened Coil

  • FESP 5133-9 - Radiating Loop_

 

The magnetic, handheld coil FESP 5133 was designed to generate defined magnetic field strength in the audio frequency range up to 3 MHz. The main application is immunity testing against magnetic fields according to  VG 95377 Part 13 and many others. Depending on the current source characteristics magnetic fields up to 400 A/m can be generated for a short time.

There are further rings available on request in order to provide a certain scaling between coil current and magnetic field strength (e.g. 1 Amp coil current = 10 A/m field strength).

 

Application as magnetic field probe:

The FESP 5133-9 can also be used to measure existing magnetic fields. The open circuit output voltage is directly proportional to the magnetic field strength (at fixed frequency) or directly proportional to frequency (at constant magnetic field strength).

 

Number of turns: 9
Wire diameter: 1.2 mm Cu
Maximum coil current: 11 A (5 min.)
Nominal coil current: 7 A continuous
Max. magn. field strength: 380.4 A/m (5 min.)
Nominal magn. field strength: 242 A/m continuous
Magnetic field strength, 1 A coilcurrent: (with 50 mm distance ring):

34.58 A/m

Current required for 1 A/m:
(with 50 mm distance ring)

28.9 mA
Flux density with 50 mA at 50 mm distance:

127 dBpT

Inner coil diameter:

133 mm
Medium coil diameter: 134.2 mm
Spacing coil center to measurement plane: 50 mm
Mechanical dimensions:  0.16 m x 0.2 m
Connector: N-jack
Usable frequency range: 10 kHz - 3 MHz
Inductance: ~ 20 μH

Resistance:

0.1 Ω

Weight:

0.8 kg

Standard:

VG 95377 Part 13

 

Datasheet FESP 5133-9

 

 


 

FESP 5132 - Radiating Loop

 
  • Schwarzbeck FESP 5132
  • Schwarzbeck FESP5132

 

The magnetic, handheld coil FESP 5132 was designed to generate defined magnetic fieldstrength in the audio frequency range up to 100 kHz. The main application is immunity testing against magnetic fields according to EN 55103 and MIL-STD-461 E. Depending on the current source characteristics magnetic fields up to 1500 A/m can be generated for a short time.

The generated magnetic fieldstrength is proportional to the coil current. A distance ring allows a precise spacing of 50 mm between coil and EUT surface. Other distance rings can be used to scale the current-fieldstrength ratio. The highest fieldstrength levels can be achieved using the shortest ring. There are further rings available on request in order to provide a certain scaling between coil current and magnetic fieldstrength (e.g. 1 Amp coil current = 100 Amp/m fieldstrength).

 

Application as magnetic field probe:

The FESP 5132 can also be used to measure existing magnetic fields. The open circuit output voltage is directly proportional to the magnetic fieldstrength (at fixed frequency) or directly proportional to frequency (at constant magnetic fieldstrength).

 

Number of turns: 20
Wire diameter: 2 mm Cu (AWG 12)
Maximum Coil Current: 20 A (5 min.)
Nominal Coil Current: 10 A continuous
Max. Magn. Field Strength: 1500 A/m (5 min.)
Nominal Magn. Field Strength: 750 A/m continuous
Magn. Field Strength, 1 A Coil Current:
(with 50 mm distance ring)

75.56 A/m
(95 µT)

Current required for 1 A/m:
(with 50 mm distance ring)

13.23 mA
Medium coil diameter: 120 mm
Spacing coil center to measurement plane: 50 mm
Mechanical Dimensions: 0.16 m x 0.25 (0.3) m
Terminals: 4 mm
Usable Frequency Range: 0 - 150 kHz

Resistance:

0.1 Ω

Weight:

1.2 kg

Standard:

ISO 11452-8
MIL 461E RS101
EN  55103 5.18.3.2

 

Datasheet FESP 5132

Security Notes FESP 5132

 

Option:

Option LoopHolder50: Calibration fixture to hold FESP 5134-40 in FESP 5132 in a distance of 50 mm acc. MIL461E figure RS101-3.

 

Datasheet LoopHolder 50


 


 

FESP 5133-1330 - Circular Coil

 
  • Schwarzbeck FESP5133_1330 circular radiating loop

 

The handheld coil FESP 5133-1330 was designed for immunity testing against magnetic fields up to approx. 20 kHz. The coil has 225 turns of Cu-wire on 15 layers. There are spacers to provide a constant distance of 50 mm between coil center and EuT-surface. The generated magnetic fieldstrength is direct proportional to the coil current. Typical Applications are Immunity Tests according to the Measuring Method SF 01 G acc. to VG 95377 standard.

 

Medium coil diameter: 126 mm
Medium Coil Distance to Surface: 50 mm
Number of turns: 225
Wire Diameter: 2 mm
Mechanical Dimensions: 190 x 190 x 290 mm
Terminals: 4 mm Laborstecker
Usable Frequency Range: 0 - 20 (50) kHz
Inductance: 6.5 mH
Resistance: 0.5 Ω
Capacitance: 240-300 pF
Resonance Frequency: 120 kHz
Impedance: |Z|=42 @ 1 kHz
Max. Current: 20 A
Max. Fieldstrength at 50 mmspacing: 17 kA/m

Max. magnetic fieldstrength at 1 A coil current in 50 mm distance:

858.2 A/m
Required Current for 1000 A/m in 50 mm distance: 1.165 A
Magnetic Flux Density at 1 A coil current in 50 mm distance:

1.078 mT
180.66 dBpT

Conversion Current-Magnetic Fielstrength: 58.67 dB/m

Weight:

5.3 kg

Standard:

Measuring Method SF 01 G acc. to VG95377

 

Datasheet FESP 5133-1330

Security Notes FESP 5133-1330

 

 


 

FESP 5135 - Circular unshielded Coil

 
  • Schwarzbeck FESP 5135

 

The magnetic loop FESP 5135 was designed to generate magnetic fields in the frequency range up to 300 kHz. The FESP 5135 is esp ecially used for magnetic immunity testing for large EuT which does not fit between Helmholtz coils. The FESP 5135 is equipped with spacers in order to achieve good repeatability of test results without using additional length measuring equipment. The magn etic fieldstrength generated by the coil is strictly proportional to the coil current, therefore a current measurement ca be used to determine the actual fieldstrength.

 

Medium coil diameter: 0.5 m
Number of turns: 20
Turn layers: 1
Mechanical Dimensions: 500 x 500 x 140 mm
Terminals: 4 mm
Usable Frequency Range: 0 - 300 kHz
Inductance: 485 µH
Resistance: 0.8 Ω
Capacitance: 45 pF
Resonance Frequency: ≈1 MHz
Magnetic Fieldstrength at 1 A coil current: 32.016 A/m
30.1 dBA/m
Continuos Current: <5 A
Maximum Current: 7 A (5 min.)
Max. Magnetic fieldstrength: 224 A/m (5 min.)
Fieldstrength (cont. operation): 160 A/m

Weight:

5.2 kg

Spacer:

100 mm

Standard:

EN 55103 - 2 A.3.1

 

Datasheet FESP 5135

 

 


 

FESP 5133 - Circular screened Coil

 
  • Schwarzbeck FESB5133 Loop Sensor  Antenna

 

The magnetic, handheld coil FESP 5133 was designed to generate defined magnetic field strength in the audio frequency range up to 100 kHz. The main application is immunity testing against magnetic fields according to e.g. (of professional studio equipment acc. to EN 55103), MIL-STD 461, VG 95377 part 13) and many others). Depending on the current source characteristics magnetic fields up to 1000 A/m can be generated for a short time.

The generated magnetic field strength is proportional to the coil current. Two different distance rings (50 mm and 100 mm separation) are supplied to ensure a well defined spacing between the coil plane and the EuT-surface. The highest field strength levels can be achieved using the shortest ring. There are further rings available on request in order to provide a certain scaling between coil current and magnetic field strength (e.g. 1 Amp coil current = 100 Amp/m field strength)

 

Application as magnetic field probe:

The FESP 5133 can also be used to measure existing magnetic fields. The open circuit output voltage is directly proportional to the magnetic field strength (at fixed frequency) or directly proportional to frequency (at constant magnetic field strength).

 

Number of turns: 36
Wire diameter: 1.26 mm Cu
Maximum coil current: 10 A (5 min.)
Nominal coil current: 5 A continuous
Max. magn. field strength: 1385 A/m (5 min.)
Nominal magn. field strength: 700 A/m continuous
Magnetic field strength, 1 A coilcurrent: (with 50 mm distance ring):

138.5 A/m

Current required for 1 A/m:
(with 50 mm distance ring)

7.22 mA
Magnetic field strength, 1 A coil current: (with 100 mm distance ring) 46.88 A/m

Current required for 1 A/m:
(with 100 mm distance ring)

21.3 mA
Flux density with 50 mA at 50 mm distance:

138.8 dBpT
136.8 dBµA/m

Inner coil diameter:

133 mm
Medium coil diameter: 135.5 mm
Spacing coil center to measurement plane: 50 mm / 100 mm
Mechanical dimensions:  0.16 m x 0.25 (0.3) m
Coil terminals: 4 mm Laborstecker (Standard) oder BNC
Shield: Electrostatic
Usable frequency range: 0 - 200 kHz
Inductance: ≈ 300 μH
Resonance frequency: >500 kHz

Resistance:

0.6 Ω

Weight:

1.1 kg

Standard:

MIL 461
EN 55103 - 2 A.2.1 b)
EN 55103 - 2 A.4.1

 

Datasheet FESP 5133

 

 


 

FESP 5210-1 - Square Induction Coil

 
  • Schwarzbeck FESP 5210-1 Square Induction Coil
  • Schwarzbeck FESP 5210-1
  • Schwarzbeck FESP5210-1

 

The square induction coil FESP 5210-1 is used to generate magnetic fields for immunity testing according to IEC 61000-4-8, IEC 61000-4-9 and IEC 61000-4-10. The edge length of the square shaped coil is 1 m. The coil comes with wing terminals, which allow a versatile and simple connection to the current source.

 

Operation as Helmholtz Coil:

Two square induction coil FESP 5210-1 can be combined to form a Helmholtz coil pair. The main advantages of a Helmholtz coil pair is an increase of fieldstrength and a significant improve- ment of field uniformity. For this purpose the coil planes are aligned in parallel. Electrically the coils have to be wired in series, keeping in mind that the current through the coils must flow in the same direction. This is very important to obtain the wanted addition of both fieldstrength contributions in the volume between the coils.

 

Frequency range: DC ... 10 MHz
Edge length: 1 m +/- 1 %
Number of turns: 1
Max. cont. current: 100 A

Max. current (10 min):

150 A
Max. current (1 min): 200 A
Max. cont. fieldstrength
(coil center):
90 A/m

Inductance:

3.4 µH
Self resonance: 16 MHz
Material / diameter: Aluminium 12 mm
Mounting tube: 22 x 195 mm
Weight: 1.5 kg
Connector: 4 mm wing terminals

Standard:

 

IEC 61000-4-8
IEC 61000-4-9
IEC 61000-4-10

 

Datasheet FESP 5210 - 1


 


 

FESP 5410-1 - Square Induction Coil

 
  • FESP 5410-1 with MFPO 9760 Current Transformer and Pulse Generator
  • FESP 5410-1
  • FESP 5410-1_Square Induction Coil

 

The square induction coil FESP 5410-1 is used to generate magnetic fields for immunity testing according to IEC 61000-4-8, IEC 61000-4-9 and IEC 61000-4-10. The side length of the square shaped coil is 1 m. The coil comes with screw terminals.

 

Operation as Helmholtz Coil:

Two square induction coil FESP 5410-1 can be combined to form a Helmholtz coil pair. The main advantages of a Helmholtz coil pair is an increase of fieldstrength and a significant improvement of field uniformity. For this purpose the coil planes are aligned in parallel. Electrically the coils have to be wired in series, keeping in mind that the current through the coils must flow in the same direction. This is very important to obtain the wanted addition of both fieldstrength contributions in the volume between the coils.

 

Frequency range: DC ... 10 MHz
Side length: 1 m +/- 1 %
Number of turns: 1
Magnetic field strength @ 1 A coil current: 0.9 A/m
Max. cont. current: 400 A
Max current (10 min): 600 A
Max. cont. field strength (coil center): 360 A/m

Inductance:

2.9 µH
Material / Cross-section: Aluminium / 10 x 25 mm
Mounting tube: 22 x 195 mm
Weight: 4.1 kg
Connector: M16 thread (male)

Standard:

 

IEC 61000-4-8
IEC 61000-4-9
IEC 61000-4-10

 

Datasheet FESP 5410 - 1


 

 

MFPO 9760 - Current Transformer and Pulse Generator

 
  • 9760_Front
  • 9760_Rueck
  • FESP 5410-1 with MFPO 9760 Current Transformer and Pulse Generator

 

The MFPO 9760 is a current transformer for the typical mains frequencies of 50 Hz / 60 Hz. It is used to feed the 1 m square induction coil FESP 5410-1. This coil can reach magnetic field strengths of 1000 A/m for short time and 360 A/m continuously in its center. As this coil has only one turn it has relatively low inductance and can be used for fast signals in the time domain. On the other hand a one turn coil has a low coil factor and needs to be driven with hundreds of Ampere to reach the fields expected by IEC 61000-4-8.

 

The pulse generator which is built into MFPO 9760 allows defined “ON intervals” of 0.5…5.0 seconds. Other time intervals can be set by manually holding the start button longer. Alternatively the interval can be set usiing the remote input.

 


Technical data MFPO 9760 with FESP 5410-1
Power requirement:  230 V / 50/60 Hz / 16 A
Max. cont. current into coil:  400 A
Max current (5 min):  600 A
Max current (max 5 sec.): 1200 A
Max. field strength (coil center, max 5 sec.): >1000 A/m
Dimensions MFPO 9760 (W x D x H): 650 x 800 x 490
Weight: approx. 60 kg
Connector: male threads M16

Standard:

IEC 61000-4-8

 

Datasheet MFPO 9760